OrthoPediatrics PediLoc Locking Plate System, 4.5mm Contour Locking Compression Femur 8 Hole Plate Right. Used for pediatric patients as indicated for pelvic, small and long bone fractures.

Company
Orthopediatrics Corp
Recall Initiated
June 14, 2012
Posted
July 20, 2012
Terminated
April 30, 2013
Recall Number
Z-2057-2012
Quantity
15 devices
Firm Location
Warsaw, IN

Reason for Recall

The device is a Left 8 hole plate but is incorrectly labeled and etched as a Right device.

Distribution

Worldwide Distribution -- USA, including states of IN, GA, MI, FL, MO, OH, and SC and country of Australia.

Lot / Code Info

Lot number: 7736703 and part number: 00-1050-4308

Root Cause

Error in labeling

Action Taken

On 15-June-2012, a follow-up, URGENT Recall of MIS-BRANDED PediLoc Locking Plate e-mail was sent to all Distributors and direct representatives to provide them with a written notification of the voluntary recall. The e-mail identified the affected product and the reason for the recall. It also instructed them to quarantine all identified devices from the affected lot number immediately. A certified Voluntary Recall notice was also sent as a follow up along with a return prepaid pouch for the return of the recalled device(s) and Mandatory Reply Form.