RecallDepth

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

This recall has been terminated (originally issued June 18, 2012).

Company
Philips Healthcare Inc.
Recall Initiated
May 11, 2012
Posted
June 18, 2012
Terminated
October 6, 2016
Recall Number
Z-1821-2012
Quantity
6
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

Distribution

6 US customers

Lot / Code Info

All systems with Eleva software version 4.x and stitching option.

Root Cause

Device Design

Action Taken

Philips sent an Urgent - Field Safety Notice letter dated May 11, 2012 to their affected customers. The letter identified the affected product, what the issues are, under what circumstances they may occur and the actions to be taken to avoid or minimize the occurrence of the issue Philips plans to supply an Addendum to the Instruction for Use. A Philips Service Engineer will contact the affected customers as soon as the Field Action Kit is ready to be implemented. For questions contact your local Philips representative: Technical Support Line 1-800-722-9377.

More recalls by Philips Healthcare Inc.

View all recalls by this company →