DYONICS POWER Shaver Blades, 4.5mm Long Full Radius, package of 3 Catalog Number: 7205341 Product Usage: Smith & Nephew DYONICS Disposable Endoscopic Blades are indicated for resection of soft and osseous tissues in both large and small articular cavities.

This recall has been terminated (originally issued July 1, 2011).

Company
Smith & Nephew, Inc. Endoscopy Division
Recall Initiated
May 6, 2011
Posted
July 1, 2011
Terminated
May 9, 2012
Recall Number
Z-2732-2011
Quantity
555 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Product has insufficient weld between the shaft tube and tip, could result in the tip separating from the shaft tube during use.

Distribution

Worldwide Distribution -- USA (nationwide) and countries of: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Germany, Hong Kong, Italy, Japan, Mexico, Netherlands, Turkey, and United Kingdom.

Lot / Code Info

Lot Number: 50561144

Root Cause

Process control

Action Taken

Smith & Nephew notified customers by an "URGENT - PRODUCT RECALL 1st NOTIFICATION" letter dated May 6, 2011. The letter describes the product, problem, and actions to be taken. The letter asked customers to call 800-343-5717 (option 3) or send an email to endo.andreturn@smith-nephew.com for a Return Authorization (RA) Number to return product, obtain credit or order replacement blades. Customers are to complete the form and return to the address provided being sure to reference the RA number. For any questions regarding this recall customers were instructed to call 508-337-3731.