Spinal Concepts polyaxial screw driver, AO, Manufactured by Spinal Concepts, Austin, TX, REF 775-01.

This recall has been terminated (originally issued December 10, 2010).

Company
Zimmer Inc.
Recall Initiated
December 1, 2005
Posted
December 10, 2010
Terminated
June 1, 2011
Recall Number
Z-0602-2011
Quantity
220
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The devices may fracture, resulting in surgical delay or the need for intervention to remove fragments from the surgical site.

Distribution

Worldwide Distribution, including US.

Lot / Code Info

Lots 12XM, 16BT, 18FY, 19GS, 20YQ, 21CB and 25ZP.

Root Cause

Device Design

Action Taken

The affected instruments were removed and replaced in the U.S. Abbott Spine and Spinal Concepts instrument kits between December 2005 and October 2007. User accounts were not advised of the removal Zimmer has confirmed that none of these instruments are currently available for use in the U.S. For further information about this action, contact Dale Miller of Zimmer at 574 372-4962.