Bio-Rad's Monolisa (TM) Anti-HBs EIA test kit, Labeled as, "MONOLISA TM Anti-HBs EIA 25220; MONOLISA TM Anti-HBs Calibrator Kit 25219... Bio-Rad Laboratories Redmond, WA 98052" For the qualitative detection of antibody to Hepatitis B Surface Antigen (anti-HBs) in Human Serum and EDTA or citrated plasma.

This recall has been terminated (originally issued May 20, 2010).

Company
Bio-Rad Laboratories
Recall Initiated
August 11, 2008
Posted
May 20, 2010
Terminated
May 20, 2010
Recall Number
Z-1620-2010
Quantity
795 kits
Firm Location
Redmond, WA
Official Source
View on FDA website ↗

Reason for Recall

The package insert provided with this kit contains a typographical error, giving incorrect directions to prepare a Working Conjugate Solution. If these directions are followed, it would result in a 1:6 dilution and invalidate the assay.

Distribution

Worldwide Distribution -- USA and Canada.

Lot / Code Info

Lot #'s: 288BEE and 056CEE

Root Cause

Packaging process control

Action Taken

All domestic customers were notified via an Urgent Product Correction customer letter sent by DHL or FedEx on 8/11/08. Enclosed with the letter was a corrected package insert. The Canadian subsidiary was notified via e-mail on 8/11/08. The letter advised customers of the correct wording, and advised that a new corrected package insert was enclosed and to make note of the correction in the laboratory procedures. Questions should be directed to Bio-Rad Laboratories Technical Support at 1-800-224-6723.