RecallDepth

Leksell GammaPlan. Leksell GammaPlan is designed for use with the Leksell Gamma Knife manufactured by Elekta Instrument AB. The Leksell GammaPlan is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgery and stereotactic radiation therapy. It processes the inputs of the health care professional (Neurosurgeons, Radiation Therapists and Radiation Physicists) such that the desired radiation dose is provided by the Leksell Gamma Knife to a precisely defined target area within the cranium.

This recall has been terminated (originally issued April 27, 2010).

Company
Elekta, Inc.
Recall Initiated
October 23, 2008
Posted
April 27, 2010
Terminated
December 14, 2010
Recall Number
Z-1134-2010
Quantity
72 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

Although the co-registration looks good during the verification step in the co-registration dialog, the obtain transformation may include an error that depends on the voxel sizes and acquisition parameters of the co-registered image studies.

Distribution

Nationwide Distribution -- (AL, AZ, AR, CA, CO, FL, GA, HI, IL,KY, MA, MI, MS, NE, NV, NV, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN TX and UT).

Lot / Code Info

4113, 4115, 4142, 4148, 4158, 4160, 4176, 4191, 4193, 4195, 4207, 4208, 4300, 4302, 4303, 4306, 4308, 4309, 4313, 4314, 4319, 4326, 4328, 4329, 4332, 4333, 4337, 4339, 4340, 4341, 4342, 4343, 4344, 4346, 4349, 4352, 4353, 4354, 4363, 4366, 4824, 4830,5000, 5003, 5019, 5023, 5026, 5034, 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6047 and 6049.

Root Cause

Other

Action Taken

Important Notice A325 "Error in ImageMerge" dated November 21, 2008 was sent to affected customers and field service engineer as notification of the problem. The notice advises users not to use ImageMerge except if it is judged that the transformation error has no clinical significance. Information is provided in the release to help the user to make an informed decision on this issue. Field Change Order 200 084, "NTPS 8.3 SP1 for NTPS 8.0, 8.2 and 8.3" released January 28, 2009 was sent to affected customers as notification of the fix to this issue. This notice includes instructions to assist Elekta field service personnel in successfully upgrading the affected systems. Once the upgrade is in place, the issue will be corrected.