RecallDepth

Synthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion, Adjustable Drill Stop, Part Number 03.613.012. Synthes Vectra System, Synthes Vectra-One System and Synthes Vectra-T System is intended for anterior plate and screw fixation of the cervical spine (C2-C7).

Company
Synthes Usa (Hq), Inc.
Recall Initiated
August 6, 2012
Posted
February 4, 2013
Terminated
August 31, 2015
Recall Number
Z-0782-2013
Quantity
48
Firm Location
West Chester, PA

Reason for Recall

Although the reported incidence is low, there is the potential for the drill stop to slip on the drill bit resulting in an adverse event.

Distribution

Nationwide Distribution-including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.

Lot / Code Info

Adjustable Drill Stop, Part Number 03.613.012 - Lots 5456424, 5494147, 5557938, and 5557939

Root Cause

Component design/selection

Action Taken

Synthes sent a "NOTICE: MEDICAL DEVICE RECALL" letter dated August 6, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 610-719-5450 for questions related to this notice.

More recalls by Synthes Usa (Hq), Inc.

View all recalls by this company →