Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion.

Company
Synthes Usa (Hq), Inc.
Recall Initiated
January 11, 2013
Posted
March 15, 2013
Terminated
August 31, 2015
Recall Number
Z-0956-2013
Quantity
4
Firm Location
West Chester, PA

Reason for Recall

It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).

Distribution

Distributed in the states of IN, OH, PA, and TX.

Lot / Code Info

Lot number 7965181

Root Cause

Labeling mix-ups

Action Taken

The firm initiated their recall of this product on January 11, 2013 by sending an Urgent Notice: Medical Device Recall letter to consignees. The letter identified the affected product and the issue. Synthes requested customers to examine their inventory for the affected product, remove them from use, and return them to Synthes. Additionally, customers were to follow the steps provided in the event that they did or did not have any affected product on hand. Questions should be directed to 610-719-5450 or FieldAction@synthes.com.