The GlideScope Video Laryngoscope (GVL) is a video laryngoscope that incorporates a high-resolution color camera, an LED light source, a rechargeable lithium battery, and video output to a remote monitor. It consists of a video laryngoscope and a portable video monitor. The affected GVL and Part numbers are GVL 3 0574-0007, GVL 4 0574-0001, and GVL 5 0574-0030. The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures.

This recall has been terminated (originally issued December 7, 2012).

Company
Verathon, Inc.
Recall Initiated
October 25, 2012
Posted
December 7, 2012
Terminated
September 27, 2013
Recall Number
Z-0478-2013
Quantity
total 2710 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Re-usable video laryngoscope blades of the GlideScope GVL3, GVL4, and GVL5 were recalled due to potential cracking.

Distribution

Worldwide distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NV, NJ, NY, OK, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY; and countries of: Argentina, Australia, Canada, China, France, Germany, Indonesia, Japan, The Netherlands, Pakistan, Panama, Peru, Philippines, Singapore, South Korea, Taiwan, Thailand, United Kingdom, and Venezuela.

Lot / Code Info

Serial numbers range for GVL3: 0574-0007/ MD105000 to MD112387; Serial numbers range for GVL4: 0574-0001/ LG105000 to LG112758; Serial numbers range for GVL5: 0574-0030/ XL105000 to XL111798.

Root Cause

Device Design

Action Taken

The firm, Verathon, sent an "URGENT - Medical Device Recall" letter dated October 25, 2012, to their consignees/cutomers. The letter described the product, problem, and actions to be taken. The customers were instructed to return the recalled blades back to Verathon and complete and return the Device Recall Notification Reply Form via fax to (425) 883-2896 or email: cservice@verathon.com. All returned products will be quarantined and will be destroyed later. Verathon will replace the affected video laryngoscope blade at no cost to customers. Customers in Canada and US with questions can call 800-331-2313. If you are located outside of the US or Canada, please call 425-867-1348.