CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.

This recall has been terminated (originally issued July 2, 2014).

Company
Nihon Kohden America Inc
Recall Initiated
June 3, 2014
Posted
July 2, 2014
Terminated
April 28, 2015
Recall Number
Z-1979-2014
Quantity
430
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.

Distribution

Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.

Lot / Code Info

Model Number: CNS-6201  Software Version 02-26 only. Serial Numbers: 00485,  00488,  00489,  00490,  00491,  00361,  00429,  00430, 00432,  00436,  00443,  00447,  00478,  00486,  00546,  00564,  00573,  00574,  00575,  00576,  00592,  00626, 00638, 00643,  00651,  00674,  00675,  00677,  00678, 00679, 00687.

Root Cause

Software design

Action Taken

Nihon Kohden sent an Urgent Medical Device Safety Notification and Field Correction letter dated June 3, 2014 to affected customers. Recall notification letter, dated June 3, 2014 was hand-delivered to site by Nihon personnel. The letter provided the following information: product description with codes and software version, reason for recall, field correction information and actions to be taken, MedWatch reporting information, and contact information for recall.