Cochlear Baha Soft tissue gauge 6mm. Reusable instrument used to ensure the soft tissue thickness is equal to or less than 6 mm. For use with Baha Attract hearing system.

This recall has been terminated (originally issued June 2, 2014).

Company
Cochlear Americas Inc.
Recall Initiated
April 9, 2014
Posted
June 2, 2014
Terminated
October 6, 2014
Recall Number
Z-1699-2014
Quantity
288
Firm Location
Centennial, CO
Official Source
View on FDA website ↗

Reason for Recall

Cochlear America is recalling Cochlear Baha Soft Tissue Gauge due to possible manufacturing imperfection in the smooth surface of the metal that could potentially leave residual metal on the patient's skin.

Distribution

Nationwide Distribution.

Lot / Code Info

Outer Package Lot COH523440, COH485876, and COH484221 with/Device Lot No. 131008

Root Cause

Process control

Action Taken

Customers were notified via letter on 4/17/14. This communication included an explanation of the recall and instructions to quarantine and return products.