Reusable Clip Applier Manufactured by AtriCure, Inc., 6217 Centre Park Drive, West Chester, OH 45069. The Reusable Clip Applier (Applier) is intended to be used to deploy the Gillinov-Cosgrove LAA Clip (Clip). The Gillinov-Cosgrove LAA Clip (Clip) is an implantable device used for occlusion of the left atrial appendage.

This recall has been terminated (originally issued March 14, 2014).

Company
Atricure Inc
Recall Initiated
January 23, 2014
Posted
March 14, 2014
Terminated
May 13, 2014
Recall Number
Z-1209-2014
Quantity
46 units
Firm Location
Chester, OH
Official Source
View on FDA website ↗

Reason for Recall

The firm was notified by a customer of a cracked spring in the handle of the RCD1 device.

Distribution

Nationwide Distribution-including the states of AL, CA, CO, CT, DE, FL, KY, MD, MI, MN, MO, NC, NY, OH, OR, PA, TN, UT, VA & WA.

Lot / Code Info

Lot Number MFR-0424-1

Root Cause

Nonconforming Material/Component

Action Taken

On 1/24/2014 the firm sent Urgent Recall Notification letters to their customers.