RecallDepth

SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

This recall has been terminated (originally issued December 31, 2013).

Company
Steris Corporation
Recall Initiated
November 21, 2013
Posted
December 31, 2013
Terminated
June 16, 2014
Recall Number
Z-0616-2014
Quantity
6,028 units
Firm Location
Mentor, OH
Official Source
View on FDA website ↗

Reason for Recall

Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.

Distribution

Nationwide Distribution - USA (nationwide) and the District of Columbia and Puerto Rico. Product was also shipped to the following countries: Afghanistan, Bahrain, Hong Kong, Italy, Saudi Arabia, Turkey & United Kingdom.

Lot / Code Info

Model #6500 & 7500

Root Cause

Device Design

Action Taken

STERIS sent an Urgent Voluntary Field Correction Notice dated November 21, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that a STERIS Field Service Representative would contact their facility to schedule installation of the new hardware and updated software on their SYSTEM IE Liquid Chemical Sterilant Processing System(s). Customers were instructed that they may continue to operate the SYSTEM IE processor prior to the completion of the update. Customers with questions were instructed to contact STERIS Field Service Dispatch at 1-800-333-8828. For questions regarding this recall call 440-392-7519.

More recalls by Steris Corporation

View all recalls by this company →