HEART CATH CUSTOM PACK; REF: DYNJ39747A; Packaged in Mexico by: Medline Industries, Inc., Mundelein, IL 60060; 1-800-Medline; www.medline.com Surgical kit used to supply tools required to perform angiography medical procedures.
- Company
- Medline Industries, Inc.
- Recall Initiated
- August 14, 2013
- Posted
- February 21, 2014
- Terminated
- December 19, 2017
- Recall Number
- Z-1048-2014
- Quantity
- 160 units
- Firm Location
- Waukegan, IL
Reason for Recall
Guidewire .035x150 3MMJ TCFC, item number 88241, has a potential for the coating to flake off of the wire. The guidewire is used in a number of surgical convenience kits assembled and marketed by Medline Industries, Inc.
Distribution
Nationwide Distribution: US to states of: AL, CA, FL, IL, LA, OH and VA.
Lot / Code Info
Pack Number: DYNJ39747A; Lots: 13EB2095, 13EB3153, 13FB4157, 13RB0832, 13GB2692
Root Cause
Nonconforming Material/Component
Action Taken
The firm, Medline Industries, Inc., sent a "RECALL IMMEDIATE ACTION REQUIRED" letter dated August 14, 2013, to its direct accounts/customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately check their stocks for inventory of the affected products, quarantine any affected products, return the product using the enclosed shipping slips for credit-If additional stickers are needed, please call 866-359-1704, and complete and return the enclosed FAX-BACK VERIFICATION FORM listing the quantity of affected product(s) on hand via fax to Medline at 866-767-1290. If you have any questions, please contact 866-359-1704.