Codman Certas Programmable Valve In Line Valve with Siphonguard, Unitized Catheter and Accessories Product Code: 82-8806 Product Usage: The CODMAN CERTAS Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular perssure drainage of CSF. The CODMAN CERTAS Therapy Management System (TMS) allows the non-invasive reading or adjustment of the valve setting.

This recall has been terminated (originally issued January 29, 2014).

Company
Codman & Shurtleff, Inc.
Recall Initiated
May 16, 2013
Posted
January 29, 2014
Terminated
June 2, 2015
Recall Number
Z-0863-2014
Quantity
409 units
Firm Location
Raynham, MA
Official Source
View on FDA website ↗

Reason for Recall

Codman Certas Programmable Valves used for hydrocephalus may not operate properly

Distribution

Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Brazil, Switzerland, Czech Republic, Germany, Spain, Finland, France, Great Britain, Israel, Italy, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa, and UK.

Lot / Code Info

Lot Codes:  CMJCMD CMBC2Z CMDB7C CMJC49 CMBCPR CMCC2T CMDBHC CMLBLZ CMMB14 CNDBCY CNDCV2 CNJCFT CNKBH6 CNLB57 CNMDNL CNNCWD CNPB1Z CPBB3J CPBDBB CPCCM0 CPDBYK

Root Cause

Device Design

Action Taken

Codman issued an Urgent Notice dated May 16, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Mitigation instructions are provided and if not followed or are not successful, either issue could lead to symptoms associated with under- or over-drainage. Units not implanted are to be returned. A U.S Acknowledgement Form and fax are requested to be completed and faxed to: 1-888-943-4193, or e-mail the form to codman8894@stericycle.com. If you have any questions on concerns in regards to this recall, please contact your Codman Representative or Stericycle at 1-888-943-2395.