Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0

This recall has been terminated (originally issued November 14, 2013).

Company
Raysearch Laboratories Ab Sveavaegen 9 Stockholm Sweden
Recall Initiated
October 23, 2013
Posted
November 14, 2013
Terminated
February 17, 2017
Recall Number
Z-0281-2014
Quantity
195 units domestically
Official Source
View on FDA website ↗

Reason for Recall

Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.

Distribution

US Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.

Lot / Code Info

Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4.

Root Cause

Software design

Action Taken

Medical Device Correction Letters dated 10/23/13 were e-mailed to the user accounts on 10/23/13.