Universal Titanium Prosthesis, Round Head, with Flex H/A titanium shoe. The ear prosthesis is implantable and indicated for hearing loss.

Company
Medtronic Xomed, Inc.
Recall Initiated
August 23, 2013
Posted
September 26, 2013
Terminated
November 7, 2013
Recall Number
Z-2284-2013
Quantity
60
Firm Location
Jacksonville, FL

Reason for Recall

One lot of the product was assembled using the Cam Head version of the same prosthesis instead of the Round head Shaft.

Distribution

Nationwide Distribution (US), including the states of CA, NE, GA, and IL.

Lot / Code Info

REF: 1150000 Lot: 0207029284

Root Cause

Process control

Action Taken

An "URGENT PRODUCT RECALL" letter was faxed to customers following the initial telephone notification. The letter described the product issue and provided recommended actions. Questions were directed to customer service at 800-874-5797.