RecallDepth

Synergetics 25 gauge Soft Tip Cannula, packaged in 6 units per box. Catalog number: 30.08.25. Single-use 25 gauge cannula used in vitreoretinal surgery.

This recall has been terminated (originally issued October 22, 2013).

Company
Synergetics Inc
Recall Initiated
July 30, 2013
Posted
October 22, 2013
Terminated
April 7, 2014
Recall Number
Z-0051-2014
Quantity
48/6-cannula boxes
Firm Location
O Fallon, MO
Official Source
View on FDA website ↗

Reason for Recall

One lot of 25 gauge Soft-Tip Disposable Cannulas are co-mingled with 23 gauge cannulas; the product is labeled as 25 gauge cannulas.

Distribution

Nationwide Distribution-USA including the states of MO, CO, KY, TN, NC, PA, IL CA, NY, IN, MI, LA, NJ and OH.

Lot / Code Info

Lot number: M222240

Root Cause

Labeling Change Control

Action Taken

The firm sent recall notification letters to their consignees on 7/30/13.

More recalls by Synergetics Inc

View all recalls by this company →