AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems intended for use in hemodialysis applications. They are designed to pre-treat and purify potable water for use in making dialysate for hemodialysis and to meet current AAMI and Federal (U.S.) standards. The AmeriWater Portable MROS model is intended for use in a hospital, clinic, dialysis center, or for home care for single patient use. The AmeriWater Portable MRO1 model is for treatment of up to two patients in a hospital, clinic, or dialysis centers.

This recall has been terminated (originally issued September 6, 2012).

Company
Ameriwater Inc
Recall Initiated
August 9, 2012
Posted
September 6, 2012
Terminated
April 17, 2014
Recall Number
Z-2321-2012
Quantity
80 Units
Firm Location
Dayton, OH
Official Source
View on FDA website ↗

Reason for Recall

The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate.

Distribution

Worldwide distribution: USA (nationwide) including states of: AL, AR, CO, CT, FL, GA, IL, IN, KS, MA, MD, MO, MS, NC, NY, OH, OK, SC, TN, TX and VA; and country of: Canada.

Lot / Code Info

Model Number MROS, Catalog #00MROS20, 00MROS21 & Model Number MRO1, Catalog #00MRO120, and 00MRO121; Model Number(s): MROS and MRO1; Serial #'s 31000 to 31079.

Root Cause

Nonconforming Material/Component

Action Taken

The firm, AmeriWater, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated August 9, 2012, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately examine their inventory and quarantine the product subject to recall; remove affected devices from service until the correction is completed to prevent malfunction of the device during patient treatment; and complete and return the attached response form to AmeriWater within 30 days via fax to 937/461-1988 Attn: Quality Assurance. AmeriWater dealer will be contacting their customers to schedule replacement of the fuses. If you have any questions, contact your local AmeriWater dealer or AmeriWater customer service at 1-800-535-5585.