Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.
This recall has been terminated (originally issued February 15, 2011).
- Company
- Elekta, Inc.
- Recall Initiated
- October 2, 2009
- Posted
- February 15, 2011
- Terminated
- February 17, 2011
- Recall Number
- Z-1251-2011
- Quantity
- 379 units
- Firm Location
- Norcross, GA
- Official Source
- View on FDA website ↗
Reason for Recall
New protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.
Distribution
Nationwide
Lot / Code Info
GFD259, GFD129, GFD729, GFD708, GND229, GND116, GFD296, GFD201, GND83, GFD815, GFD2127, GFD2128, GFD2130, GFD2131, GND148, GFD06, GND44, GFD621, GFD352, GND157, GND181, GFD312, GFD106, GFD351, GND79, GFD1000, GFD1001, GFD249, GFD2247, GFD2265, GFD874, GND227, GND55, GFD529, GFD770, 9903005129, GND223, GFD216, GFD416, GFD915, GFD908, SH00208, SH00226, GFD688,M GFD746, GND246, GFD526, GFD571, GFD976, GFD633, GFD879, GFD919, GFD1013, GFD841, GFD869, GFD979, GFD12, GFD808, GFD524, GFD940, GFD271, GFD517, GFD760, GFD327, GND176, GFD780, GFD779, GFD412, GFD634, GFD42, GFD700, GND184, GFD538, GND76, GFD280, GND39, GFD109, GFD337, GND208, GFD65, GFD481, GFD212, GFD870, GFD792, GND24, GFD817, GFD776, GFD81, GND30, GFD163, GFD21, GND45, GND62, GFD1164, GFD861, GFD862, GFD1053, GFD1057, SH00008, SH0013, SH0033, GFD2316, GFD2318, GND147, GFD938, GFD657, GFD920, GFD924, GND109, GFD18, GFD494, GFD495, GND230, GFD559, GFD707, GFD692, GFD966, GFD968, GFD165, GFD133, GFD522, GFD520, GFD523, GFD762, GFD821, GND258, GFD43, GFD0054, GFD0055, GND178, GND36, GFD358, GFD256, GND80, GFD1076, GFD138, GFD617, GFD184, GFD194, GFD130, GFD96, GFD2174, GFD2272, GFD1062, GFD140, GFD117, GFD264, GND65, 9906005540, GFD804, GND165, GFD823, GFD1060, GFD101, GFD189, GND60, GFD137, GND94, GFD1066, GND216, GFD689, GFD223, GFD83, GFD99, GFD381, GFD706, FD662, GFD664, GFD237, GND185, GFD845, GFD26, GND126, GFD277, GFD284, GFD2245, GFD2246, GND135, GFD50, GFD200, GFD228, GFD2270, GFD2221, GND233, GND264, GFD671, GFD343, GFD91, GFD652, GFD281, GFD414, GND231, GND71, GFD120, GFD582, GFD690, GFD696, GFD697, GFD878, GFD71, GFD360, GFD368, GND40, GFD17, GFD658, GFD152, GFD660, GFD893, GND200, GFD831, GFD850, GFD69, GND122, GFD840, GFD698, GFD709, GND48, GND152, GND52, GFD1080, GFD809, GFD1059, GFD2264, GFD2229, GND192, GFD145, GFD820, GFD198, GFD188, GND172, GFD822, GND16, GND57, GFD41, GFD912, GFD917, GFD267, GFD998, GFD247, GFD255, GND10, GND123, GND31, GFD68, GFD632, GFD74, GFD934, GND159, GFD239, GFD261, GFD235, GFD521, GFD613, GFD399, GFD161, GND47, GFD843, GFD863, GFD865, SHO0165, GFD190, GFD816, GFD328, GFD557, GFD558, GND190, GFD19, 000000000000002, GFD824, GND27, GFD618, GND226, GND35, GND134, GND41, GND238, GFD257, GFD132, GFD543, 9904005345, GFD555, GFD143, GFD1000, GFD300, GND215, GND108, GFD586, GFD288, GND250, GFD033, GFD218, GND221, GND224, GFD107, GFD832, GFD278, GFD295, GFD1005, GFD725, GFD573, GFD667, GFD680, GFD956, GND194, GFD1971, GFD630, GFD587, GND28, SH00159, SH00160, GFD361, GFD370, GFD619, GFD620, GFD384, GFD156, GFD362, GFD914, GFD301, GFD1054, GND263, GFD733, GFD735, GFD913, GND189, GND167, GFD909, GFD376, GFD116, GFD903, GFD699, GFD317, GND161, GFD906, GFD186, GND193, GFD58, GFD234, GFD519, GND89, GFD730, GND261, GFD527, GFD391, GFD250, GFD763, GFD103, GFD483, GFD51, GND63, GND86, GFD778, GFD82, GFD948, GFD382, GFD747, GFD825, GFD556, GND75, GFD480, 9902005057, GND245, GND191, GND269, GND111, GFD205, GND84, GND170, GFD67, GFD232, GND96, GND100, GND22, GFD554, GND49, GND17, GFD2249, GFD2260, GFD1946
Root Cause
Labeling design
Action Taken
Elekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009. The Important Notice will be added to the Customer User manual upon receipt.