Elekta Leksell Stereotactic System, used for localization (spatial reference) for cranial surgery using X-ray or CT and MRI Image data.

This recall has been terminated (originally issued February 15, 2011).

Company
Elekta, Inc.
Recall Initiated
October 2, 2009
Posted
February 15, 2011
Terminated
February 17, 2011
Recall Number
Z-1251-2011
Quantity
379 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

New protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.

Distribution

Nationwide

Lot / Code Info

GFD259, GFD129, GFD729, GFD708, GND229, GND116, GFD296, GFD201, GND83, GFD815, GFD2127, GFD2128, GFD2130, GFD2131, GND148, GFD06, GND44, GFD621, GFD352, GND157, GND181, GFD312, GFD106, GFD351, GND79, GFD1000, GFD1001, GFD249, GFD2247, GFD2265, GFD874, GND227, GND55, GFD529, GFD770, 9903005129, GND223, GFD216, GFD416, GFD915, GFD908, SH00208, SH00226, GFD688,M GFD746, GND246, GFD526, GFD571, GFD976, GFD633, GFD879, GFD919, GFD1013, GFD841, GFD869, GFD979, GFD12, GFD808, GFD524, GFD940, GFD271, GFD517, GFD760, GFD327, GND176, GFD780, GFD779, GFD412, GFD634, GFD42, GFD700, GND184, GFD538, GND76, GFD280, GND39, GFD109, GFD337, GND208, GFD65, GFD481, GFD212, GFD870, GFD792, GND24, GFD817, GFD776, GFD81, GND30, GFD163, GFD21, GND45, GND62, GFD1164, GFD861, GFD862, GFD1053, GFD1057, SH00008, SH0013, SH0033, GFD2316, GFD2318, GND147, GFD938, GFD657, GFD920, GFD924, GND109, GFD18, GFD494, GFD495, GND230, GFD559, GFD707, GFD692, GFD966, GFD968, GFD165, GFD133, GFD522, GFD520, GFD523, GFD762, GFD821, GND258, GFD43, GFD0054, GFD0055, GND178, GND36, GFD358, GFD256, GND80, GFD1076, GFD138, GFD617, GFD184, GFD194, GFD130, GFD96, GFD2174, GFD2272, GFD1062, GFD140, GFD117, GFD264, GND65, 9906005540, GFD804, GND165, GFD823, GFD1060, GFD101, GFD189, GND60, GFD137, GND94, GFD1066, GND216, GFD689, GFD223, GFD83, GFD99, GFD381, GFD706, FD662, GFD664, GFD237, GND185, GFD845, GFD26, GND126, GFD277, GFD284, GFD2245, GFD2246, GND135, GFD50, GFD200, GFD228, GFD2270, GFD2221, GND233, GND264, GFD671, GFD343, GFD91, GFD652, GFD281, GFD414, GND231, GND71, GFD120, GFD582, GFD690, GFD696, GFD697, GFD878, GFD71, GFD360, GFD368, GND40, GFD17, GFD658, GFD152, GFD660, GFD893, GND200, GFD831, GFD850, GFD69, GND122, GFD840, GFD698, GFD709, GND48, GND152, GND52, GFD1080, GFD809, GFD1059, GFD2264, GFD2229, GND192, GFD145, GFD820, GFD198, GFD188, GND172, GFD822, GND16, GND57, GFD41, GFD912, GFD917, GFD267, GFD998, GFD247, GFD255, GND10, GND123, GND31, GFD68, GFD632, GFD74, GFD934, GND159, GFD239, GFD261, GFD235, GFD521, GFD613, GFD399, GFD161, GND47, GFD843, GFD863, GFD865, SHO0165, GFD190, GFD816, GFD328, GFD557, GFD558, GND190, GFD19, 000000000000002, GFD824, GND27, GFD618, GND226, GND35, GND134, GND41, GND238, GFD257, GFD132, GFD543, 9904005345, GFD555, GFD143, GFD1000, GFD300, GND215, GND108, GFD586, GFD288, GND250, GFD033, GFD218, GND221, GND224, GFD107, GFD832, GFD278, GFD295, GFD1005, GFD725, GFD573, GFD667, GFD680, GFD956, GND194, GFD1971, GFD630, GFD587, GND28, SH00159, SH00160, GFD361, GFD370, GFD619, GFD620, GFD384, GFD156, GFD362, GFD914, GFD301, GFD1054, GND263, GFD733, GFD735, GFD913, GND189, GND167, GFD909, GFD376, GFD116, GFD903, GFD699, GFD317, GND161, GFD906, GFD186, GND193, GFD58, GFD234, GFD519, GND89, GFD730, GND261, GFD527, GFD391, GFD250, GFD763, GFD103, GFD483, GFD51, GND63, GND86, GFD778, GFD82, GFD948, GFD382, GFD747, GFD825, GFD556, GND75, GFD480, 9902005057, GND245, GND191, GND269, GND111, GFD205, GND84, GND170, GFD67, GFD232, GND96, GND100, GND22, GFD554, GND49, GND17, GFD2249, GFD2260, GFD1946

Root Cause

Labeling design

Action Taken

Elekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009. The Important Notice will be added to the Customer User manual upon receipt.