DRIVER STAR #8 CANNULATED DART-FIRE (TM), COMPRESSION SCREW, SIZE: #8, REF: DSDS-0008, 1 EACH, Rx ONLY, NON-STERILE, EC REP Wright Medical EMEA, Krijgsman 11, 1186 DM Amstelveen, The Netherlands, Wright Medical Technology, Inc., 5677 Airline Road, Arlington, TN 38002

This recall has been terminated (originally issued September 22, 2010).

Company
Wright Medical Technology Inc
Recall Initiated
August 26, 2010
Posted
September 22, 2010
Terminated
August 2, 2012
Recall Number
Z-2470-2010
Quantity
13 units
Firm Location
Arlington, TN
Official Source
View on FDA website ↗

Reason for Recall

After autoclaving, the epoxy identifier (colored) band bled and there is concern that it may flake.

Distribution

TX, PA, TN, VA, GA, and Spain

Lot / Code Info

Lot number 040754491

Root Cause

Process design

Action Taken

The firm initiated their recall the product by letter on 08/26/2010. The letters were addressed to distributors and hospital administrators with a fax back response form. Please contact Customer Service at 800-238-7117, if you have any questions.