CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is indicated for the treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

This recall has been terminated (originally issued January 11, 2012).

Company
Accuray Incorporated
Recall Initiated
November 7, 2011
Posted
January 11, 2012
Terminated
February 27, 2012
Recall Number
Z-0382-2012
Quantity
30
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.

Distribution

Worldwide Distribution -- USA ( nationwide ) including the states of FL, TN, IL, NJ, CO, PA, MD, CA, MS, MN, SC, WA, OK, GA, AZ, and TX., and the countries of France and United Kingdom.

Lot / Code Info

Model number: PN 032384

Root Cause

Labeling design

Action Taken

Accuray sent a Urgent Medical Device Correction letter dated November 7, 2011, via Federal Express to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The information provided in this letter will be incorporated in user manuals shipped with the CyberKnife system. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at 1. (877) 668-8667 ( USA) 1. (408) 716-4700 ( non USA )