OR table TITAN UT, Surgical Table. Surgical bed with Orthopedic Accessories.

Company
Trumpf Medical Systems, Inc.
Recall Initiated
January 21, 2011
Posted
January 11, 2012
Terminated
August 13, 2012
Recall Number
Z-0398-2012
Quantity
see line 1.
Firm Location
Charleston, SC

Reason for Recall

Product is being recalled due to unintended movement on the Titan Tables.

Distribution

Worldwide Distribution.

Lot / Code Info

Material number 1465166, All Serial numbers.

Root Cause

Device Design

Action Taken

Trumpf Medical Systems sent an initial notification to customers dated January 24, 2011. A subsequent "URGENT SAFETY NOTICE-UPDATE" letter dated May 26, 2011 was sent to all affected customers. The letter described the product, problem, and actions to be taken by the customers. A Corrective Action Plan includes a software update to be installed by a Trumpf Technical Service Representative. Contact Technical Service at 888-474-9359 for questions regarding this notice.