Lap-Band Access Port Needle 10-Pack, 20 G x 51 mm (2 in.), Product Code: 8-20302-10, Lot Number: 9L01N 20 gauge, 51 mm (2 in.) length non-coring deflected tip (Huber) needle, used with the LAP-BAND System for band adjustment.

This recall has been terminated (originally issued May 2, 2011).

Company
Allergan
Recall Initiated
March 21, 2011
Posted
May 2, 2011
Terminated
April 6, 2012
Recall Number
Z-2123-2011
Quantity
1878 x 10-packs
Firm Location
Goleta, CA
Official Source
View on FDA website ↗

Reason for Recall

The recall was initiated because Allergan has determined that a number of Access Port Needle pouches from one specific lot number may have small gaps in a pouch seal. Consequently the product may no longer be sterile.

Distribution

Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, MN, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Australia, Canada, Germany, Great Britian, and the Netherlands

Lot / Code Info

Lot Number: 9L01 N

Root Cause

Pending

Action Taken

Allegan sent an Urgent Medical Device Recall letter dated March 21, 2011, to all affected customers. Allergan requested that upon receipt of letter, that the customer check their existing inventory of product for affected lot. Customers were requested to complete Business Reply Card, stating quantity of inventory of affected lot, if any, and return to Allergan Medical. Customer with remaining inventory were instructed to contact Allergan Medical Product Support at 1-800-624-4261, x5972 or 1-805-961-5972 for return and immediate replacement. . Customers will be issued credit for returned product. For questions regarding this recall call 805-319-0589.

More recalls by Allergan