BD Vacutainer Push Button collection Set; Non-Pyrogenic, Sterile, Rx only; 23 G x 3/4 x 12" (0.6 x 19 mm x 305 mm) Becton Dickinson & Co., Franklin Lakes, NJ, USA Intended use: Blood collection

This recall has been terminated (originally issued December 27, 2010).

Company
Becton Dickinson & Company
Recall Initiated
October 5, 2010
Posted
December 27, 2010
Terminated
June 6, 2012
Recall Number
Z-0816-2011
Quantity
4,223,200
Firm Location
Franklin Lakes, NJ
Official Source
View on FDA website ↗

Reason for Recall

There is the possibility for certain lots of the BD Vacutainer Push Button Blood Collection sets to have separation of the cannula from the hub.

Distribution

Nationwide Distribution

Lot / Code Info

Catalog Number 367342; Lot numbers: 0145457, 0152300, 0153981, 0153982, 0158081, 0158082, 0158084, 0160664, 0160666, all exp 5/31/2012; Lot numbers 0165202, 0165205, 0165207, 0165218, 0165226, 0165227, 0165232, 0165234, 0176872, 0179208, 0179209, 0179212, 0180023, 0180026, 0180027, 0180028, 0180040, 0180042, all exp 6/30/2012; Lot numbers 0196209, 0196210, 0201399, 0201400, 0201402, 0208593, 0211641, all exp 7/31/2012.

Root Cause

Process control

Action Taken

BD Diagnostics sent "RECALL NOTIFICATION" separate letters October 1 and 5, 2010 and follow up letter dated October 8, 2010, as well as a letter dated December 1, 2010 to all distributors and customers commencing October 5, 2010 by federal express. The letters describe the product, the problem and the action to be taken by the customers. The customers were instructed to immediately examine their inventory for specific catalog and lot numbers, and return product from the lots following the instructions on the enclosed packing slip (BD will send replacement upon receipt of returned product). If you have any questions on returning and replacing product, please contact BD logistics at 1-866-380-0046. If you have any further detail questions regarding this recall, please contact your BD Sales Consultant or BD Technical Services via email at www.BD.com/Vacutainer/Contact or via telephone at 1-800-631-0174.