Leksell Gamma Knife Perfexion, Product Number: 715000. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

This recall has been terminated (originally issued April 5, 2010).

Company
Elekta, Inc.
Recall Initiated
November 6, 2009
Posted
April 5, 2010
Terminated
December 14, 2010
Recall Number
Z-1232-2010
Quantity
32 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

There may be a situation where the Frame Adapter might lock the stereotactic Frame in the wrong position.

Distribution

Nationwide Distribution -- AZ, CA, FL, IL, KY, MI, MN, MS, MO, NY, NC, OH, OR, PA, SC, TX, UT and VA.

Lot / Code Info

Serial Numbers: 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6045, 6047, 6048, 6049, 6053, 6054, 6056, 6057, 6059 and 6061.

Root Cause

Software design

Action Taken

Elekta, Inc. issued a Field Change Order 200 091 titled "Update Frame Adapter with New Frame Fixation Axle" dated November 6, 2009. Consignees were given instructions to prevent further issues related to the affected device. For further information, contact Elekta, Inc. at 1-770-300-9725.