Synthes Zero-P StarDrive Screwdriver Shaft, T8, self-retaining, quick coupling, for use in spinal fusion surgery. Catalog # 03.617.902 Technique Guide.

This recall has been terminated (originally issued June 10, 2009).

Company
Synthes Spine
Recall Initiated
February 10, 2009
Posted
June 10, 2009
Terminated
October 5, 2009
Recall Number
Z-1348-2009
Quantity
1766 units
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

Tip breakage, due to inadequate instructions for use. The Technique Guide provided incorrect information related to use of the Torque Limiting Attachment during screw insertion and final tightening.

Distribution

Distributed Nationwide and Canada.

Lot / Code Info

Catalog # 03.617.902 - All lots

Root Cause

Labeling design

Action Taken

All affected Synthes Spine Sales Consultants were notified of the Labeling Correction via email on February 10, 2009. They were instructed to review changes with their surgeons. They were supplied with an electronic version of the corrected Technique Guide on March 11, 2009. All affected Synthes Sales Consultants and Customers were notified of this labeling change via distribution of the updated Technique Guide and a notification letter. This packet was sent via USPS certified mail on March 26, 2009. To ensure a broad and accurate distribution of notification, affected consultants and customers are defined as having been shipped Zero-P 3.0 mm Titanium Cervical Spine Locking Screws. These devices are used only with the Zero-P system.