Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog Number 801046, Terumo Cardiovascular Systems, Ann Arbor, MI 48103

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
September 10, 2007
Posted
December 1, 2007
Terminated
July 7, 2009
Recall Number
Z-0323-2008
Quantity
120
Firm Location
Ann Arbor, MI

Reason for Recall

Failure to operate: The unit display may go blank and reset when the start button is pressed, due to an internal electrical short, resulting in a centrifugal pump that fails to start as expected.

Distribution

Nationwide.

Lot / Code Info

Serial Numbers 0020 through 0041, 0043, 0044, 0046 through 0060, 0062 through 0091, 0094, 0098, 0118 through 0137, 0139 through 0156, 0159 and 0166 through 0176.

Root Cause

Component change control

Action Taken

Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, given instructions to follow if the problem were to manifest itself, and informed the firm would contact them to schedule an upgrade.