Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.

This recall has been terminated (originally issued February 15, 2007).

Company
Stille-Sonesta, Inc
Recall Initiated
November 9, 2006
Posted
February 15, 2007
Terminated
June 20, 2007
Recall Number
Z-0504-2007
Quantity
66 units.
Firm Location
Dallas, TX
Official Source
View on FDA website ↗

Reason for Recall

Bolts may back out of their joints causing the back section of the table to tilt.

Distribution

Nationwide.

Lot / Code Info

Serial Numbers:  12-70000 - 12-70017, 00010, 00012, 00013 - 00022, 00024 - 00027, 00030 - 00034, 00036 - 00040, 00043 - 00044, 00045 - 00049, 00053 - 00054, 00057 - 00065, 00067 - 00069, 00071.  Outside U.S. Serial Numbers: 12-70018 - 12-70019, 23-70000 - 23-70009, 23-70008, 23-700010, 00011, 00023, 00028 - 00029, 00050 - 00052, 00055 - 00056, 00066, 00072, 00073, 00074.

Root Cause

Other

Action Taken

Notifications letters sent to consignees on 11/20/06. Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary. Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.