Clinac (linear accelerator); Model 21EXS Silhouette Machine (slim version of the Clinac); Distributed by and/or Manufactured by: Varian Medical Systems, Palo Alto, CA 94304.

Company
Varian Medical Systems Inc 3100 Hansen Way M/S E-210 Palo Alto Ca 94304
Recall Initiated
March 30, 2005
Posted
September 9, 2005
Terminated
September 8, 2005
Recall Number
Z-1530-05
Quantity
24 units

Reason for Recall

Emergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated.

Distribution

US distribution is: CA, CT, IL, IN, KY, OH, MI, NC, NJ, NY, SC, TN, TX, VA, & WA, totaling 22 hospitals. Foreign distribution includes: Denmark, Hong Kong, and Japan. It did not involve Defense Supply Center, VA or other Federal Government sales/distribution centers.

Lot / Code Info

Affected Serial Numbers: H140001, H140002, H140003, H140004, H140005, H140006, H140007, H140009, H140022, H140010, H14001, H140012, H140013, H140014, H140015, H140016, H140017, H140018, H140019, H140021, H140023, H140024, H140025.

Root Cause

Device Design

Action Taken

On 3/16/05, the firm initiated the recall and their notification was via letters informing its customers of the corrective action.