Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc., Covington, GA 30014, USA.

Company
C.R. Bard, Inc., Urological Division
Recall Initiated
April 2, 2009
Posted
June 29, 2009
Terminated
December 4, 2009
Recall Number
Z-1539-2009
Quantity
16,100 units
Firm Location
Covington, GA

Reason for Recall

The affected units may contain extension sets with the microbore tubing instead of extension sets with the macrobore tubing.

Distribution

Nationwide Distribution - including states of AZ, CA, CO, GA, IA, MN, NE, and TX.

Lot / Code Info

Lot #JUSJF441.

Root Cause

Process control

Action Taken

All consignees were notified by an Urgent: Product Recall Notification letter via Fed-Ex carrier on 04/02/2009. They were instructed not to use or further distribute any of the affected products. They were advised to check their inventory and to remove and return affected product to Bard Medical Division. A Recall & Effectiveness Check Form was included to be filled out and returned via fax, to Bard.