Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc., Covington, GA 30014, USA.
- Recall Initiated
- April 2, 2009
- Posted
- June 29, 2009
- Terminated
- December 4, 2009
- Recall Number
- Z-1539-2009
- Quantity
- 16,100 units
- Firm Location
- Covington, GA
Reason for Recall
The affected units may contain extension sets with the microbore tubing instead of extension sets with the macrobore tubing.
Distribution
Nationwide Distribution - including states of AZ, CA, CO, GA, IA, MN, NE, and TX.
Lot / Code Info
Lot #JUSJF441.
Root Cause
Process control
Action Taken
All consignees were notified by an Urgent: Product Recall Notification letter via Fed-Ex carrier on 04/02/2009. They were instructed not to use or further distribute any of the affected products. They were advised to check their inventory and to remove and return affected product to Bard Medical Division. A Recall & Effectiveness Check Form was included to be filled out and returned via fax, to Bard.