Boston Scientific Mach 1 peripheral Guide Catheter 7F, MP Multipurpose, .081 in (internal dia), Shaft lenght 55cm, Catalog Number: 19-723, UPN M001197230, Sterile EO, Availmed S.A. de C.V. (Mexico), Boston Scientific Corp., Maple Grove, MN 55311
- Recall Initiated
- February 7, 2008
- Posted
- May 7, 2008
- Terminated
- November 3, 2008
- Recall Number
- Z-1304-2008
- Quantity
- 10
- Firm Location
- Scientific Place Natick, MA
Reason for Recall
Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length.
Distribution
Worldwide Distribution including USA states of PA, AZ, NY, KY, CA, WV, OH, WA, MI, TN, IN, IL, NJ, KY, FL, AR, FL, NC, CO, TX, WV, WI, OR, AL, MN and ND and countries of BE, FR, GB, DE, IE, IT, JP, AU, SOUTH KOREA, NZ, TR, TW and ZA.
Lot / Code Info
Lot Number: 20030666
Root Cause
Labeling mix-ups
Action Taken
Consignees were sent a Boston Scientific Urgent Voluntary Medical Device Recall letter on 2/7/08. The letter included the customer steps for recall, a complete listing of all product batches being recalled, the Reply Verification Tracking Form(s) and a return shipping label.