Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm).
- Company
- Steris Corp
- Recall Initiated
- March 22, 2007
- Posted
- August 23, 2007
- Terminated
- September 28, 2010
- Recall Number
- Z-1193-2007
- Quantity
- 78
- Firm Location
- Mentor, OH
Reason for Recall
The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard.
Distribution
Worldwide, including USA, Denmark, Belgium, Greece, India, and Russia.There are no Canadian accounts.
Lot / Code Info
Prevac Steam Sterilizer, Model #'s V-136H, V-148H, V-160H, SV-148H, V-1264 & V-1263& SFPP Steam Sterilizer, Model #'s P-148H & P-160H. Code # B093934202, B093924204, B09324223, B093924225, B093924227, B093924229, B093924233, B093924244, B093924246, B093924248, B093924250, B093924286, B093924288, B093924296, B093924307, B093924309, B093924310, B093924434, B093924476, B093927301, B093927302, B093927322, B300903936
Root Cause
Employee error
Action Taken
By telephone 3/22/2007, customers were instructed to cease using the product and a service technician would be dispatched immediately to repair the unit.