Smiths CADD Medication Cassette Reservoir, With clamp and female Luer, Nonvented stopper included, 50 ml PC 4694-24A, REF 21-7001-24, Sterile, Made in Mexico Smith Medical MD, Inc. St Paul, MN 55112 USA

This recall has been terminated (originally issued August 30, 2007).

Company
Smiths Medical Md, Inc.
Recall Initiated
June 18, 2007
Posted
August 30, 2007
Terminated
February 10, 2011
Recall Number
Z-1214-2007
Quantity
32,100
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Smith Medical has become aware that an increase in pH may occur with some medications when instilled in the Affected Cassettes. Also, a small number of catheter occlusions have been reported from one customer when these cassettes have been used for delivering Flolan.

Distribution

Worldwide, including USA, Canada, Mexico, Sweden, Germany, Japan, Australia, Taiwan, Poland, Jordan, Czech Republic, Denmark, Paraguay, Hong Kong, Singapore, Argentina, Italy, Venezuela, Greece, Norway, Pakistan, Puerto Rico, Thailand, United Arab Emirates, Netherlands Antilles, Malaysia, Panama, Hungary, Saudi Arabia, Portugal, South Africa, Denmark, the Netherlands, France, United Kingdom, Great Britain, Spain, Finland, Guam, Virgin Islands, China, Israel, Austria, and Switzerland.

Lot / Code Info

102X16, 117X16, 118X16, 130X16, 133X16, 135X16, 139X16, 147X16, 150X16, 158X16, 165X16, 178X16, 180X16, 181X16, 188X16, 191X16, 194X16, 202X16, 205X16, 226X16

Root Cause

Other

Action Taken

All customers who have received the CADD Medication Cassette reservoirs have been sent the Product Recall Notification Letter on June 18, 2007 making them aware of the potential issue. Customers will be required to return the affected devices to Smiths Medical. All returned recall affected product will be quarantined and then destroyed. The consignees are to fill and fax "Attachment 1" form in regard to affected product inventory to Smiths Medical MD.