RecallDepth

OEC FluoroTrak 9800 PLUS, image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT.

This recall has been terminated (originally issued September 28, 2007).

Company
Ge Oec Medical Systems, Inc
Recall Initiated
March 1, 2007
Posted
September 28, 2007
Terminated
May 1, 2012
Recall Number
Z-0680-2007
Quantity
5 units
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

During routine service, primary and/or secondary collimators may not have been properly reinstalled.

Distribution

Nationwide and Germany.

Lot / Code Info

Serial numbers: 89-1579-N, 89-2529-N, 82-2160-N, 8S-1932-N, 8S-2226-N.

Root Cause

Other

Action Taken

Customers were notified by letter on 03/01/2007. They were told to continue using the machines and that an inspection would be scheduled immediately to verify that collimators had been correctly installed.

More recalls by Ge Oec Medical Systems, Inc

View all recalls by this company →