Hill-Rom Affinity Three Birthing Bed, Model P3700.

This recall has been terminated (originally issued September 14, 2006).

Company
Hill-Rom, Inc.
Recall Initiated
August 18, 2006
Posted
September 14, 2006
Terminated
July 14, 2009
Recall Number
Z-1503-06
Quantity
9,324
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission of smoke.

Distribution

Nationwide, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, France, French Polynesia, Germany, Guam, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Sait Kitts and Nevis, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, U.S. Virgin Islands and Venezuela.

Lot / Code Info

All units distributed after October 2000 and equipped with the 12VAC accessory outlet option.

Root Cause

Other

Action Taken

Customers were notified via letter dated 8/18/06 and instructed to either correct the problem themselves with the instructions provided or to contact the firm requesting a visit from the recalling firm to correct their beds.