Solara 2G manual wheelchair.
- Company
- Invacare Corporation
- Recall Initiated
- June 27, 2006
- Posted
- September 20, 2006
- Terminated
- October 11, 2007
- Recall Number
- Z-1509-06
- Quantity
- 36 wheelchairs
- Firm Location
- Elyria, OH
Reason for Recall
The wheelchairs with the Travel Ready Option (TRRO) may contain incorrectly finished oval tie down brackets. The inner edge of the bracket may cause the tie-down straps (used to tie down the chairs during transport) to fray or be cut.
Distribution
Nationwide and Canada.
Lot / Code Info
06EE008858, 06EE010422, 06EE010432, 06EE010434, 06EE010442, 06EE010453, 06EE010454, 06EE010745, 06EE010747, 06EE010749, 06EE010758, 06EE010799, 06EE010804, 06EE010810, 06EE011357, 06EE011378, 06EE011379, 06FE000048, 06FE000050, 06FE000054, 06FE000055, 06FE000057, 06FE000985, 06FE001000, 06FE001012, 06FE001604, 06FE001612, 06FE001625, 06FE006654, 06FE006661, 06FE007243, 06FE008820, 06FE008829, 06FE008832, 06FE009212, 06FE010083.
Root Cause
Other
Action Taken
The recalling firm notified their consignees in a letter dated 6/26/06.