SYNCHRON Clinical Systems Lipase (LIPA) Reagent; Classification name: 862.1465

This recall has been terminated (originally issued November 30, 2004).

Company
Beckman Coulter Inc.
Recall Initiated
October 25, 2004
Posted
November 30, 2004
Terminated
June 17, 2005
Recall Number
Z-0262-5
Quantity
6,698
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

Beckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.

Distribution

US and Canada

Lot / Code Info

Part number: 465101  Lot Numbers - Date of Mfg - Expiration Date M310237 - 10/29/2003 - 10/31/2004 M401010 - 01/15/2004 - 01/31/2005  M403026 - 03/05/2004 - 03/31/2005 M406041 - 06/10/2004 - 06/30/2005

Root Cause

Other

Action Taken

The week of 10/25/2004, a product corrective action (PCA) letter was mailed vis U.S. mail to all customers shipped LIPA reagent for the 24 months preceding the discovery of the issue. Customers instructed to repeat any sample that produces a lipase value of > 170 U/L that cannot be confirmed by patient history.