AngioTouch Kit Model H1000P Hand Controller kits used with the ACIST Contrast Management System.

This recall has been terminated (originally issued July 20, 2004).

Company
Acist Medical Systems
Recall Initiated
May 19, 2004
Posted
July 20, 2004
Terminated
November 3, 2005
Recall Number
Z-1002-04
Quantity
3590 kits
Firm Location
Eden Prairie, MN
Official Source
View on FDA website ↗

Reason for Recall

Some of the sterile product packages were inadequately sealed prior to sterilization and as a result the package contents may not be sterile.

Lot / Code Info

Lot # 0584B

Root Cause

Other

Action Taken

Each consignee was contacted by telephone and asked to return unused kits with the affected lot number.