Hill-Rom Affinity Three Birthing Bed, Model P3700A.

This recall has been terminated (originally issued February 22, 2008).

Company
Hill-Rom, Inc.
Recall Initiated
November 21, 2007
Posted
February 22, 2008
Terminated
September 22, 2009
Recall Number
Z-0583-2008
Quantity
32,379 for all model numbers
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated.

Distribution

Worldwide Distribution-- including USA and countries of Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Morocco, The Netherlands, New Calidonia, New Zealand, Nigeria, Northern Mariana Islands, Norway, Oman, Panama, Philippines, Poland, Portugal, Qatar, Republic of Korea, Saudi Arabia, Singapore, South Korea, Romania, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguey, Venezuela, and Virgin Islands .

Lot / Code Info

All units. (**Affinity beds manufactured with serial number I290AA7395 or greater are not impacted by this correction.)

Root Cause

Device Design

Action Taken

Consignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com.