Intuitive brand da Vinci S Surgical System, Endoscopic Instrument Control System, Model IS1200 with Isolation Transformer (P/N: 952012) and Vision Cart (Model VS1000), Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA

Company
Intuitive Surgical, Inc.
Recall Initiated
October 4, 2007
Posted
February 22, 2008
Terminated
March 31, 2011
Recall Number
Z-0151-2008
Quantity
63 total units
Firm Location
Sunnyvale, CA

Reason for Recall

Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use of an under-rated fuse will result in failure of the fuse and a loss of power to the vision cart, including the loss of power to any ancillary equipment that m

Distribution

Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, LA, MA, MD, MN, NC, NJ, NV, OH, PA, PR, SD, TN, TX, WA, and WI, and countries of Belgium, Ireland, Italy, Korea, Russia, and the United Kingdom.

Lot / Code Info

SG015, SG053, SG060, SG067, SG096, SG126, SG151, SG164, SG201, SG245, SG267, SG269, SG270, SG271, SG272, SG273, SG275, SG276, SG278, SG280, SG282, SG283, SG285, SG286, SG287, SG288, SG289, SG290, SG291, SG292, SG293, SG294, SG295, SG296, SG297, SG298, SG299, SG300, SG301, SG302, SG303, SG309, SG312, SG313, SG314, SG315, SG316, SG317, SG318, SG319, SG323, SG324, SG325, SG326, SG329, and SG330

Root Cause

Mixed-up of materials/components

Action Taken

A product recall was initiated on October 4, 2007. Direct and Distributor consignees were sent an Important Product Notice, via electronic mail on October 5, 2007. The firm plans on sending a Field Service Engineer to inspect the system and make any necessary corrections.