Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.
- Recall Initiated
- April 1, 2004
- Posted
- June 30, 2007
- Terminated
- June 11, 2008
- Recall Number
- Z-0923-2007
- Quantity
- 525 total for both models
- Firm Location
- Ann Arbor, MI
Reason for Recall
A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease.
Distribution
Worldwide: USA, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Singapore and United Arab Emirates.
Lot / Code Info
Serial numbers 0031 through 0324 are potentially affected by the problem. Serial numbers 0092, 0128, 0152, 0223, 0226, 0237, 0246, 0260 and 0296 have been corrected.
Root Cause
Other
Action Taken
The firm is visiting consignees and making corrections as complaints are received. These corrections, to date, have been made between 04/04 and 09/06. The firm will correct other units as complaints are received and issued a Device Correction Letter dated 6/6 /07 notifying all users of the problem and appropriate responses should it occur.