Hill-Rom VersaCare Hospital Bed System; Models P3200 and P3201, Hill-Rom, Batesville, IN 47006

This recall has been terminated (originally issued July 24, 2007).

Company
Hill-Rom, Inc.
Recall Initiated
November 12, 2006
Posted
July 24, 2007
Terminated
October 22, 2010
Recall Number
Z-0240-2008
Quantity
43,731 beds total for both models were distributed; 3,282 have been corrected to date.
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

False Latch/Failure to Latch-The side rail may not latch, or the latch may not fully engage, resulting in the siderail being able to be lowered without pulling on the release handle.

Distribution

Worldwide, including USA, Argentina, Australia, Brazil, Canada, Dominican Republic, Emirate of Sharjah, England, Greece, Indonesia, Japan, Jordan, Kuwait, Lebanon, Malaysia, Mexico, Oman, Pakistan, Philippines, Qatar, Saudi Arabia, Selangor, Singapore, South Korea, Spain, Taiwan, Thailand, Turkey, United Arab Emirates and Venezuela.

Lot / Code Info

Serial numbers E322AD0046 through H023AD6179. (E means 2003 production, F means 2004 production, G means 2005 production and H means 2006 production).

Root Cause

Device Design

Action Taken

The firm notified consignees via an Urgent Device Correction letter dated 11/12/07 informing them of the problem and informed them they would be receiving upgrade kits which they could have their qualified personnel install themselves following video instructions also provided, or they could contact Hill-Rom for assistance.