RecallDepth

Ossur Total Knee 1900, Model Number TK 1900, external prosthetic knee.

This recall has been terminated (originally issued July 4, 2006).

Company
Ossur North America Inc
Recall Initiated
March 29, 2006
Posted
July 4, 2006
Terminated
August 8, 2008
Recall Number
Z-1198-06
Quantity
778
Firm Location
Aliso Viejo, CA
Official Source
View on FDA website ↗

Reason for Recall

The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising the knee function.

Distribution

Distribution worldwide in Austria, Canada, Columbia, Hong Kong, Japan, Korea, Kuwait, Malaysia, Mexico, Taiwan, & New Zealand. Involved states are: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, AND WY.

Lot / Code Info

Serial numbers: 14466 ¿ 16053 including the following repaired units outside the stated range. 12588, 13131, 13216, 13236, 13265, 13363, 13487, 13573, 13598, 13799, 13967, 14028, 14032, 14127, 14143, 14158, 14159, 14370.

Root Cause

Other

Action Taken

On 3/29/06 a press release was posted on OandP.com, an industry specific web site. An announcement listing involved serial numbers and contact phone numbers was also posted on the company''s web site, ossur.com. A two stage mailing has been constructed. The first letter sent to all customers March 30, 2006, describes the problem and models affected. The second letter, March 31, 2006, is customer specific and lists all of the involved serial numbers by model which were purchased by that customer. Telephone calls will be placed to customers who do not respond to the mailings.