RecallDepth

Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.

This recall has been terminated (originally issued December 1, 2005).

Company
Stryker Howmedica Osteonics Corp.
Recall Initiated
November 22, 2004
Posted
December 1, 2005
Terminated
October 3, 2006
Recall Number
Z-0241-06
Quantity
19
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

The Gamma Nail may have been manufactured with an incorrect lag screw hole angulation. Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees. Catalog number 3220-0320S, lot code K949408 was manufactured with a lag screw hole angulation of 125 degrees instead of 120 degrees.

Distribution

The product was sent to Stryker Orthopaedics branches and agents throughout the US. There was one hospital that was a direct account.

Lot / Code Info

Catalog 3220-0320S Lot K949408  3225-0360S, lot code K949396

Root Cause

Other

Action Taken

Notification letters and Product Accountability forms were sent via FedEx on 11/22/2004 with return receipt.