COULTER LH 750 Hematology Analyzer.

This recall has been terminated (originally issued February 10, 2005).

Company
Beckman Coulter Inc.
Recall Initiated
December 16, 2004
Posted
February 10, 2005
Terminated
April 26, 2012
Recall Number
Z-0536-05
Quantity
1,257
Firm Location
Brea, CA
Official Source
View on FDA website ↗

Reason for Recall

A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.

Distribution

US and Canada

Lot / Code Info

Part Number 6605632

Root Cause

Other

Action Taken

Letter 12/20/04 A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation''s database are not affected. If the LH750 instrument displays a pop-up winder titled Database Error (XXXXXX) and text similar to (Microsoft) (ODBC SQL...) As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Do not acknowledge the error. Instead, restart the entire system by completely powering it OFF and then ON. Or, if the LH750 Workstation experiences a change in performance, such as screen display freezes, slow response time, or requires frequent resets to continue processing. As soon as possible, call Beckman Coulter Customer Service, 800-526-7694) or contact your local rep, to schedule a service call. Verify the accuracy of all patient reports for the samples until service confirms the performance of system.

More recalls by Beckman Coulter Inc.

View all recalls by this company →