Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit

This recall has been terminated (originally issued April 6, 2005).

Company
Diasorin Inc.
Recall Initiated
April 12, 2004
Posted
April 6, 2005
Terminated
July 22, 2006
Recall Number
Z-0673-05
Quantity
8
Firm Location
Stillwater, MN
Official Source
View on FDA website ↗

Reason for Recall

The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.

Distribution

CA, CT, KS, MA & WA

Lot / Code Info

Software version 12

Root Cause

Other

Action Taken

A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.