Magnetom Trio System. Nuclear Magnetic Resonance Imaging Diagnostic Device
- Recall Initiated
- December 18, 2003
- Posted
- February 10, 2004
- Terminated
- November 9, 2004
- Recall Number
- Z-0521-04
- Quantity
- 16 units
- Firm Location
- Malvern, PA
Reason for Recall
tissue roll can become magnetic
Distribution
The product was shipped to medical facilities in CA, CT, DC, FL, GA, IL, IN, KS, KY, MA, MD, MN, NC, NY, OH, OR, PA, TX, and UT.
Lot / Code Info
Model #7387074 - Serial #''s: 20501 through 20508, 20510 through 20524, and 20550 through 20570. Model #7106441 - Serial # 20500
Root Cause
Other
Action Taken
The recalling firm has issued Updated Instructions to replace the stainless steel tissue roll holder. The recalling firm''s representatives are visiting each account to replace the tissue roll holder.