RecallDepth

Thoratec brand TLC-II Portable Ventricular Assist Device (VAD) Driver; Catalog No. 20010-2085-000 Distributed and Manufactured by Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, CA 94588

Company
Thoratec Corp
Recall Initiated
August 25, 2006
Posted
October 3, 2006
Terminated
November 27, 2007
Recall Number
Z-0004-2007
Quantity
114 devices (57 valves/2 per unit)
Firm Location
Pleasanton, CA

Reason for Recall

Sticky Valve Disk- The valve disk has an increased tendency to stick during operation causing noise and pressure/vacuum alarms.

Distribution

Worldwide distribution-USA including states of AL, CA, CT, FL, IL, IN, KY, LA, MA, ME, MI, MN, NC, NY, OH, PA, TN, TX, VA, WA, and WI., with 19 Foreign consignees Australia, Belgium, Canada*, Czech Republic, France, Germany, Ireland, Netherlands, Portugal, Switzerland, United Kingdom. (*Hospital Royal Victoria, Queen Elizabeth II/Capital Health, St. Paul's Hospital, Vancouver, Canada.)--

Lot / Code Info

All Codes (Valve Disk Lot 20052904)

Root Cause

Other

Action Taken

On 8/26/06, the firm issued letters to all its direct consignees, informing them of the affected product and providing 'IMPORTANT PRODUCT SAFETY BULLETIN'.

More recalls by Thoratec Corp