Microlab AT, part numbers 135001, 135001OOC, 135002, 135003, 135005, 135006, 135006R, 135007, 135008, 135016, 135028, 135029, 936400, 936400BLU. Microlab AT+ and Microlab AT+2, part numbers 142000, 142001, 142001R, 142002, 142003, 142005, 142006, 142006R, 142007, 142008, 142009, 142015, 142017, 142035, 142036
This recall has been terminated (originally issued February 14, 2006).
- Company
- Hamilton Co
- Recall Initiated
- December 13, 2005
- Posted
- February 14, 2006
- Terminated
- May 7, 2008
- Recall Number
- Z-0514-06
- Quantity
- 1071 units
- Firm Location
- Reno, NV
- Official Source
- View on FDA website ↗
Reason for Recall
Heat sinking on power boards may not operate as intended and the board may cease to operate.
Distribution
Product was released for distribution to 132 consignees in US, and 17 International consignees
Lot / Code Info
All codes included.
Root Cause
Other
Action Taken
On 12/15/05, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.